High Temp. Transportable blackbody reference source
R1500T – High Temperature, Transportable Furnace
LANDCAL R1500T is a high temperature, transportable blackbody radiation source designed for calibration of a range of thermometers such as System 4 and Fibroptic thermometers. R1500T is a reference standard source giving a resolution of ±1°C/2°F
The LANDCAL R1500T is a high stability, transportable, blackbody, reference source designed for on-site or laboratory calibration of infrared radiation thermometers up to temperatures of 1500 °C/2750 °F.
Six silicon carbide elements heat the conical ended cylindrical blackbody cavity to 1450 °C/2650 °F in approximately 30 minutes. A 3-term controller holds the set temperature to within ±1K/2°F.
The R1500T can be used as a transfer standard, providing calibration by the comparison method, using optional standard radiation thermometers. Alternatively a traceable calibration certificate can be supplied for the source, as an optional extra, where direct traceability is required.
| Temperatures | |||
|---|---|---|---|
| Low Temperature (C): | 500 | High Temperature (C): | 1500 |
| Low Temperature (F): | 932 | High Temperature (F): | 2732 |
Technical Details
| Maximum temperature: | 1500 °C/2750 °F |
| Recommended temp.: | 500 to 1500 °C/950 to 2750 °F |
| Heating rate: | 30 min. to 1450 °C/2650 °F |
| Stability: | <±1K/2 °F over 30 minutes at set temperature |
| Radiation cavity Type: Diameter: External aperture: | Silicon carbide, 120° cone 45mm/1.8in dia x 100mm/4.0in 40mm/1.6in dia. |
| Emissivity: | Approx 0.99 at short wavelengths |
| Heating elements (6): | SiC 151/356/20/25.4/7.4 |
| Control thermocouple: | Pt 13% Rh/Pt, Type R |
| Controller: | Eurotherm with RS232C serial interface |
| Power requirement: | 110/120V a.c. or 220/240V a.c., 50 to 60Hz |
| Power consumption: | 3.0kVA |
| Overall Dimensions: | 500 x 380 x 540mm/19.7 x15.0 x 21.3in (L x W x H) |
| Weight: | 26kg/57.3lb(nett)/32kg/70.5lb (gross) |
| Uncertainty (400 to 1500°C): | ±3K/6°F (with traceable certificate) |